Common Earth/Gnd Points
In order to ensure that there are sufficient points to terminate sensor sheathing when amoured cable is used to connect a sensor to the VibWire-108 the following terminal points are internally wired in common:
Earth
Earth
Earth
Earth
Gnd
On the power supply connection terminals
Therm -
Sense -
on both the sensor input and analogue output terminals.
Fig 14 shows how the channels are wired together to form common earth connection. Effectively all of the Earth, Gnd, Sense - and Therm - terminals are wired together.
Any earth Sheathing from armoured cable etc.. can be connected to any of terminals mentioned above for ease of installation.
Lightening Protection
The lightening protection within the VibWire-108 cannot protect the instrument from a direct lightening strike. It is used to protect the instrument from local ground strikes close to the sensors and cabling.
All of the sensor inputs are protected by transorb and gas discharge tubes. The transorb are high capacitance devices and are not used on all systems as they can distort low level signals to a point where the instrument can not be accurately measured. The transorb does protect the instrument at lower levels than the gas discharge tube, and starts to become active around 12V.
The gas discharge tube protection activates at around 92V DC and resets instantaneously after
Fig 14 shows the VibWire-108 connected to a system earth using the Earth terminators mounted adjacent to the power connectors.